Abstract
In order to determine the organic acids and monomeric sugars in fermentation broth with the corn-husk hydrolysate as carbon source, aminex HPX-87H ion exclusion column was used with refractive index detector. The mobile phase was 0.005 mol/L of sulfuric acid ( pH 2. 0 ) with flow râte of 0. 6 mL/min at 55 °C. Organic acids and monomeric sugars in samples were completely separated within 16 min. AU the correlation coefficients( r) of regression equations were above 0.999 , and the recovery ranged from 99.3% to 102.1 % . The relative Standard deviations were in the range of 0.6% -1.6% . By this method, organic acids and monomeric sugars can be easily, rapidly and effectively determined.
Original language | English |
---|---|
Pages (from-to) | 605-608 |
Number of pages | 4 |
Journal | Chinese Journal of Analytical Chemistry |
Volume | 37 |
Issue number | 4 |
State | Published - Apr 2009 |
Keywords
- Cornhusk hydrolysate
- High performance liquid chromatography
- Monomeric sugars
- Organic acids
- Succinic acid fermentation broth