3D Printed Microfluidic Device with Microporous Mn2O3-Modified Screen Printed Electrode for Real-Time Determination of Heavy Metal Ions

Ying Hong, Meiyan Wu, Guangwei Chan, Ziyang Dai, Yizhou Zhang, Guosong Chen, Xiaochen Dong

科研成果: 期刊稿件文章同行评审

107 引用 (Scopus)

摘要

Fabricating portable devices for the determination of heavy metal ions is an ongoing challenge. Here, a 3D printing approach was adopted to fabricate a microfluidic electrochemical sensor with the desired shape in which the model for velocity profiles in microfluidic cells was built and optimized by the finite element method (FEM). The electrode in the microfluidic cell was a flexible screen-printed electrode (SPE) modified with porous Mn2O3 derived from manganese containing metal-organic framework (Mn-MOF). The microfluidic device presented superior electrochemical detection properties toward heavy metal ions. The calibration curves at the modified SPE for Cd(II) and Pb(II) covered two linear ranges varying from 0.5 to 8 and 10 to 100 μg L-1, respectively. The limits of detection were estimated to be 0.5 μg L-1 for Cd(II) and 0.2 μg L-1 for Pb(II), which were accordingly about 6 and 50 times lower than the guideline values proposed by the World Health Organization. Furthermore, the microfluidic device was connected to iPad via a USB to enable real-time household applications. Additionally, the sensing system exhibited a better stability and reproducibility compared with traditional detecting system which offered a promising prospect for the detection of heavy metal ions especially in household and resource-limited occasions.

源语言英语
页(从-至)32940-32947
页数8
期刊ACS Applied Materials and Interfaces
8
48
DOI
出版状态已出版 - 7 12月 2016

指纹

探究 '3D Printed Microfluidic Device with Microporous Mn2O3-Modified Screen Printed Electrode for Real-Time Determination of Heavy Metal Ions' 的科研主题。它们共同构成独一无二的指纹。

引用此