摘要
A simple method to amplify the electrochemical signal by an aptamer with 22 bases modified with CdS hollow nanospheres (CdSHNs) was described. Using the thrombin as a model, the interaction between the aptamer and CdSHNs was characterized by cyclic voltammetry, electrochemical impedance spectroscopy and circular dichroism spectroscopy. CdSHNs promoted the electron transfer between the gold electrode and K3[Fe(CN)6] and facilitated the conformation conversion of the aptamer from hairpin to G-quadruplex after the aptamer interacted with thrombin. Under optimal conditions, the modified electrode could be used for the determination of thrombin from 0 to 33μgmL-1 and the sensitivity was 1.34μAmLμg-1cm-2, while the linear range of the modified electrode without the immobilization of CdSHNs was from 2.75 to 27.5μgmL-1 and the sensitivity was 0.062μAmLμg-1cm-2. This constructed biosensor also had a good stability, specificity, reproducibility and accuracy which could provide a promising platform for fabrication of aptamer based biosensors.
源语言 | 英语 |
---|---|
页(从-至) | 3531-3535 |
页数 | 5 |
期刊 | Biosensors and Bioelectronics |
卷 | 26 |
期 | 8 |
DOI | |
出版状态 | 已出版 - 15 4月 2011 |
已对外发布 | 是 |