Determination of osimertinib mesylate by HPLC

Ping Zhou, Li Li, Lulu Wu, Chenhao Cu, Huimin Tian, Xiaoming Ren, Haijun Zhang, Jiaquan Wu

科研成果: 期刊稿件文章同行评审

3 引用 (Scopus)

摘要

To establish an HPLC method to determine osimertinib mesylate, Agilent ZORBAX Eclipse Plus Clg column (4.6 mm x 250 mm, 5 jjliii) was used with a mobile phase consisting of methanol-buffer solution (20 mmol/L NaH2P04, pH 3.0 adjusted with 85% H3P04) (50 : 50) at the flow rate of 1.0 mL/min. The detection wavelength was 210 nm, and the column temperature was kept at 35 The calibration curve was liner over the range from 50% to 150% of determination concentration (0. 201 1-0. 603 2 mg/mL, r =0. 999 9). The limit of quantitation ( LOQ) and limit of detection ( LOD) were 0. 32 p.g/mL and 0. 08 fig/mL, respectively. The contents of osimertinib mesylate in samples were 100. 1%,99. 5% and 99. 7%. Good chromatographic separation of osimertinib mesylate and its related substances, including synthetic impurities and degradation products, were obtained. The established HPLC method is specific, accurate, simple and durable, and could be used for the determination of osimertinib mesylate.

源语言英语
页(从-至)322-327
页数6
期刊Journal of China Pharmaceutical University
48
3
DOI
出版状态已出版 - 6月 2017