Electrochemical sensor based on a silver nanowires modified electrode for the determination of cholesterol

Lin Xu, Yiting Hou, Mengdan Zhang, Tao Cheng, Wei Huang, Cheng Yao, Qiong Wu

科研成果: 期刊稿件文章同行评审

19 引用 (Scopus)

摘要

In this study, the fabrication of an amperometric cholesterol biosensor based on silver nanowires (AgNWs) and cholesterol oxidase (ChOx)-graphene oxide (GO)-chitosan (CS) film is reported. The electrochemical behaviour of the ChOx/Ag/GO/CS/ITO biosensor was studied using cyclic voltammetry (CV), which revealed that the developed biosensor possessed high sensitivity (13.628 μA mM-1 cm-2) and a low detection limit (0.427 mg dL-1). The apparent Michaelis-Menten constant, KappM, of this biosensor was very low (2.813 mM), originating from the effective immobilization process and the nanowires structure of the substrate. The biosensor expressed a wide linear range up to 400 mg dL-1 in physiological conditions (pH 7.0), making it very promising for the clinical determination of cholesterol.

源语言英语
页(从-至)5649-5653
页数5
期刊Analytical Methods
7
13
DOI
出版状态已出版 - 7 7月 2015

指纹

探究 'Electrochemical sensor based on a silver nanowires modified electrode for the determination of cholesterol' 的科研主题。它们共同构成独一无二的指纹。

引用此